Replace the Multi Purpose Valve

Tools required
p/n
5023-2502
Parts required
#
1
1
1
2
1
Remove all capillary connections from the Multi Purpose
Valve.
3
Unscrew the black union nut and remove the head of the
purge valve by pulling it to the front.
1260 Infinity II Flexible Pump User Manual
Description
Hex driver SW-6.35, slitted
p/n
Description
0100-1259
Blank nut (plastic)
01080-83202
Blank nut (stainless steel)
5067-4237
Multi Purpose Valve Head
5500-1253
Capillary ST 0.17 mm x 130 mm SX/S (OPTIONAL)

Replace the Multi Purpose Valve

2
Remove the clamp with the inline filter (if installed).
4
Put the new valve head onto the valve drive such that
the lobe fits to the groove. Screw the valve head onto
the valve drive using the union nut.
7
Maintenance
165