Replace The Module Firmware - Agilent Technologies 1290 Infinity II 2D-LC User Manual

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

10
Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic

Replace the Module Firmware

When
The installation of newer firmware might be necessary
The installation of older firmware might be necessary
Tools required
Description
Agilent Lab Advisor software
Parts required
#
1
Preparations
Read the following:
Procedure to follow:
Do not mix firmware files from different firmware sets.
NOTE
2D-LC User Guide
if a newer version solves problems of older versions or
to keep all systems on the same (validated) revision.
to keep all systems on the same (validated) revision or
if a new module with newer firmware is added to a system or
if third party control software requires a special version.
Description
Firmware, tools and documentation from Agilent web site
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
292