3. Substitute a thru for the device and perform a response calibration by pressing:
Reconnect your test device.
5. To better view the measurement trace, press:
Notice that in Figure 2-34 the SAW filter under test has considerable phase shift within only
a 2 MHz span. Other filters may require a wider frequency span to see the effects of phase
shift.
The linearly changing phase is due to the device's electrical length. You can measure this
changing phase by adding electrical length (electrical delay) to compensate for it.
6. To place a marker at the center of the band, press:
7. To activate the electrical delay function, press:
.MARKER DELAY
~~~~~~~,
This function calculates and adds in the appropriate electrical delay by taking a &lo%
span about the marker, measuring the A4, and computing the delay as the negative of
Figure 2-34. Linearly changing Phase
Making Measurements 241