Error Messages in Alphabetical Order
217
auxiliary channel.
ABORTING COPY OUTPUT
Information
This message is displayed briefly if you have pressed m to abort a copy
operation. If the message is not subsequently replaced by error message
Message
number 25, PRINT ABORTED, the copy device may be hung. Press (Local) once
more to exit the abort process and verify the status of the copy device. At this
point, the copy device will probably have an error condition which must be
ADDITIONAL STANDARDSNEEDED
Error Number Error correction for the selected calibration class cannot be computed until you
6 8
have measured all the necessary standards.
Error Number ybu have sent a read command to the analyzer (such as ENTER 716) without
31
The analyzer has no data in the output queue to satisfy the request.
AIR FLOW RESTRICTED: CHECK FAN FILTER
Error Number Something is restricting the air flow into the analyzer. Check for any debris and
2 0
clean or replace the fan filter.
ALL RRGISTBRS HAVE BEEN USED
Error Number You have used all of the available registers; you can store no more instrument
2 0 0
states even though you may still have sufficient memory. There are 31 registers
available, plus the present instrument state.
1 O-2
Error Messages