6. To measure the standard, when the displayed trace has settled, press:
The
analyzer displays
The analyzer underlines the @!?Zl& softkey after it measures the standard.
7. Disconnect the open, and connect a short circuit to PORT 1.
8. To measure the device, when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the .$~@B?- softkey
9. Disconnect the short, and connect an impedance-matched load to PORT 1.
10. To measure the standard, when the displayed trace has settled, press:
The analyzer measures the load and underlines the '@@ softkey.
11. Repeat the open-short-load measurements described above,.. but connect the devices in turn
Include any adapters that you would include in your device measurement.
12. To compute the reflection correction coefficients, press:
Make a "thru" connection between the points where you will connect your device under
test as shown in F'igure 5-7.
Note
Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.
WAIT -, MEXWFlING CAL STANDARD
during the standard measurement.