HP 8753E User Manual page 589

Network analyzer
Table of Contents

Advertisement

BLOCK INPUTLENGTHEHHOR
Error Number The length of the header received by the analyzer did not agree with the size
of the internal array block. Refer to the
35
analyzer input commands.
CALIBRATION ABORTED
Error Number You have changed the active channel during a calibration so the calibration in
progress was terminated. Make sure the appropriate channel is active and
74
restart the calibration.
Error Number A calibration set could not be found that matched the current stimulus state or
63
measurement parameter. You will have to perform a new calibration.
Error Number You have attempted to initialize a floppy disk to DOS format on an external
185
disk drive that does not support writing to all 80 tracks of the double density
and high density disks. The older single-sided disks had only 66 tracks and
some disk drives were limited to accessing that number of tracks. lb format
the disk, either choose another external disk drive or use the analyzer's
internal disk drive.
Error Number You have attempted to rename, delete, or otherwise alter the factory preset
199
state. The factory preset state is permanently stored in CMOS memory and
cannot be altered. If your intent was to create a user preset state, you must
create a new instrument state, save it, and then rename it to "UPRESET".
Refer to Chapter 12, "Preset State and Memory Allocation," for more detailed
instructions.
Error Number The disk is being accessed by the analyzer and is found to contain an HFS
203
(hierarchical file system) or fles nested within subdirectories. The analyzer
does not support HP8 Replace the disk medium with a LlF or DOS formatted
disk that does not contain Eles nested within subdirectories.
HP 8753E Network Analyzer
1 O-6
Error Messages