Information The selected power range does not support the power level of one or more
segments in the swept list table. This message appears when swept list mode is
Message
not on and reports the llrst segment that is out of range. Change the segment
power or change the power range.
Information A segment entered in the swept list table caused one or more frequency
Message
segments to overlap. This message appears when swept list mode is not on and
reports the first segment that is overlapping another. Change the frequency
ranges of the overlapping segments.
Error Number The sequence you attempted to run does not contain instnunent commands
124
Service Error Internal test #n has failed. Several internal test routines are executed at
Number 112 instrument preset. The analyzer reports the iirst failure detected. Refer to the
HP 8753E Network Amlgm Semrice
internal tests and the self-diagnose feature.
Error Number The sequence running was stopped prematurely when you pressed the llocalJ
key.
157
Error Number When you pause a sequence, you cannot continue it if you have modified it.
You must start the sequence again.
153
Error Messages