Phase or Group Delay Measurements

For information on group delay principles, refer to "Group Delay Principles" in Chapter 6.
The accuracy of this measurement depends on the quality of the mixer that is being used for
calibration and how well this mixer has been characterized. The following measurement must
be performed with a broadband calibration mixer that has a known group delay. The following
table lists the specifications of two mixers that may be used for calibration:
1.
Set the LO source to the desired CW frequency and power level. For this example the LO
source is set to the following values:
CW frequency = 1000 MHz
power = 13 dRm
2.
Initialize the analyzer by pressing Q%+.
3.
From the front panel of the HP 8753E, set the desired receiver frequency and source
output power by pressing:
4.
Connect the instruments as shown in Figure 3-17, placing a broadband "calibration" mixer
Caution
To prevent connector damage, use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.
3-24
Making Mixer Measurements
Model
ANZAC
dc to 1250 MHz
Group

Hide quick links: