Figure 3-10. Connections for Broad Band Power Meter Calibration
4. Select the HP 8753E as the system controller:
5. Set the power meter's address:
8. Perform a
sweep power meter calibration over the IF frequency range at 0 dBm:
one
Because power meter calibration requires a longer sweep time, you may want
Note
to reduce the number of points before pressing ~~~,,:~~~~~~. After the
power meter calibration is finished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.
N E T W O R K
A N A L Y Z E R
I I
P O W E R M E T E R
Making Mixer Measurements 3-13