Error Number You can only nest sequences to a maximum level of six. The sequence will
164
abort if you nest more than six.
Error Number You can have a maximum of 30 segments or 1632 points in frequency list mode.
50
In power meter calibrations, you can have a maximum of 12 segments for
power sensor cal factors and power loss functions.
Error Number You can perform a time domain transformation only in linear and CW sweep
16
types.
Service Error Your equipment setup for the adjustment procedure in progress is not correct.
Number 115 Check the setup diagram and instructions
Information
In single sweep mode, the instrument ensures that aI.l changes to the
Message
instrument state, if any, have been implemented before taking the sweep. The
command that you have initiated is being processed and wiII not be complete
until the new sweep is completed. An asterisk * is displayed in the left margin
until a complete fresh sweep has been taken.
Information
This message is displayed between the start and finish of a read or write
operation to a disk.
Message