Error Number All the memory in the sequence you are modifying is filled with instrument
132
commands.
Error Number When you use a sliding load (in a user-dellned calibration kit), you must set at
71
least three slide positions to complete the calibration.
6 9
in progress. Start a new calibration.
Error Number You have no disk in the current disk unit. Insert a disk, or check the disk unit
41
number stored in the analyzer.
Service Error The self-diagnose function of the instrument operates on an internal test
Number 114 failure. At this time, no failure has been detected.
Error Number No flies of the type created by an analyzer store operation were found on the
4 5
disk or the disk drive is empty. If you requested a specific file title, that llle
was not found on the disk.
Error Number The first IF signal was not detected during pretune. Check the front panel R
5
channel jumper. If there is no visible problem with the jumper, refer to the
HP 8753E Network Anulgzer
Error Messages lo-16