Modifying TRM Standards

In order to use the TRL technique, the calibration standards characteristics must be entered
into the analyzer's user deflned calibration kit.
This example procedure shows you how to dehne a calibration kit to utilize a set of TRM
(THRU, REFLECT, MATCH) standards. This example TRM kit contains the foliowing:
zero length THRU
n
"flush" short for the REFLECT standard (0 second offset)
n
50 ohm termination for the MATCH (inhnlte length line)
n
Note
Hewlett-Packard strongly recommends that you read product note 8510-8A
before you attempt to modify the standard de5nitions The part number of this
product note is 5091-3645E. Although the product note was written for the
HP 8510 family of network analyzers, it also applies to the HP 8753E.
For a discussion on TRL calibration, refer to "TRL/LRM Calibration" in
Chapter 6, "Application and Operation Concepts. n
Modify the Standard DelMtions
1. Press the following keys to start modifying the standard delinitions:
2. To select a short, press (iJ (ZJ. (In this example the REFLECT standard is a SHORT)
3. Press the following keys: