Source Attenuator Switch Protection; Allowing Repetitive Switching Of The Attenuator - HP 8753E User Manual

Network analyzer
Table of Contents

Advertisement

Source Attenuator Switch Protection

The programmable step attenuator of the source can be switched between port 1 and port 2
when the test port power is uncoupled, or between channel 1 and channel 2 when the channel
power is uncoupled. To avoid premature wear of the attenuator, measurement configurations
requiring continuous switching between different power ranges are not allowed.
For example, channels 1 and 2 of the analyzer are decoupled, power levels in two different
ranges are selected for each channel, and dual channel display is engaged. To prevent
continuous switching between the two power ranges, the analyzer automatically engages
the test set hold mode after measuring both channels once. The active channel continues
to be updated each sweep while the inactive channel is placed in the hold mode. (The
status annotation tsH appears on the left side of the display.) If averaging is on, the test
set hold mode does not engage until the
protection feature.

Allowing Repetitive Switching of the Attenuator

. ../.....
measurements which demand repetitive switching of the step attenuator. Use these softkeys
with caution; repetitive switching can cause premature wearing of the attenuator.
occur before activating the test set hold mode.
Application and Operation Concepts
specified number of sweeps is completed. The

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents