HP 8753E User Manual page 588

Network analyzer
Table of Contents

Advertisement

ASCII: MISSING 'VAR' STATEMENT
Error Number The CITblle you just downloaded over the HP-IB or via disk was not properly
196
organized. The analyzer is unable to read the "VAR" statement.
AVERAGING INVALID ON NON-RATIO MEASURE
Error Number You cannot use sweep-to-sweep averaging in single-input measurements.
13
Sweep-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B,
and S-parameters). You can use noise reduction techniques, such as narrower
IF bandwidth, for single input measurements
BAD FREQ FOR HARMONIC OR FREQ OFFSET
Error Number You turned on time domain or recalled a calibration that resulted in start and
stop frequencies that are beyond the allowable limits.
181
BATTERY FAILED. STATE MEMORY CLEARED
Error Number The battery protection of the non-volatile CMOS memory has failed. The CMOS
memory has been cleared. Refer to the
183
Memory Allocation," for more information about the CMOS memory.
Error Number The battery protection of the non-volatile CMOS memory is in danger of failing.
If this occurs, all of the instrument state registers stored in CMOS memory will
184
be lost. Save these states to a disk and refer to the HP
State and Memory Allocation," for more information about the CMOS memory.
BLOCK INPUT ERROR
Error Number The analyzer did not receive a complete data transmission. This is usually
34
caused by an interruption of the bus transaction. Clear by pressing the ILocal
key or aborting the I/O process at the controller.
1 O-4
Error Messages
HP 8753E Network Analgzer semrice
8753E Network Anulgzer