HP 8753E User Manual page 657

Network analyzer
Table of Contents

Advertisement

Example 2, An 8510 Display Memory File
Example 2 shows a simple file that contains no frequency information. Some instruments do
not keep frequency information for display memory data, so this information is not included in
the CITIfile package.
Note that instrument-specific information (#NA = Network Analyzer information) is also stored
in this fle. This convention allows the designer to define keywords that are particular to his or
her particular implementation.
Example
CITIFILE A.O1.OO
#NA VERSION HP8510B.05.00
NAME MEMORY
VAR FREQ MAG 5
DATA S RI
BEGIN
END
Example 3,861O Data file
Example 3 shows a CITI6ie package created from the data register of an 8510 Network
Analyzer. in this case 10 points of real and imaginary data was stored, and frequency
information was recorded in a segment list table.
Example
CITIFILE A.O1.OO
NAME DATA
#NA REGISTER 1
VAR FREQ MAG 10
DATA S[l,l] RI
SEG,LIST-BEGIN
SEG 1000000000 4000000000 10
SEG,LIST-END
BEGIN
END