HP 8753E User Manual page 598

Network analyzer
Table of Contents

Advertisement

Error Number There is not enough memory space for the power meter calibration array.
154
Increase the available memory by clearing one or more save/recall registers, or
by reducing the number of points.
Error Number You pressed an undefined softkey.
2
Error Number Limit lines cannot be turned on unless a limit table has been created. Refer to
205
create a limit table.
Error Number List mode has been turned off in the frequency offset mode because it is
182
incompatible with your selected Lo frequency.
Error Number The frequency list is empty. To implement list frequency mode, add segments
9
to the list table.
Error Number A logarithmic sweep is only valid if the stop frequency is greater than four
150
times the start frequency. For frequency spans of less than two octaves, the
sweep type automatically reverts to linear sweep.
Information
The frequency domain data points must be harmonically related from dc to the
stop frequency. That is, stop =
Message
condition is not true when a low pass mode (step or impulse) is selected and
transform is turned on, the analyzer resets the start and stop frequencies. The
stop frequency is set close to the entered stop frequency, and the start
frequency is set equal to stop/n.
1 O-14
Error Messages
n
x start, where n = number of points If this