4

Plan 1X ERGO Objective

Using the Plan 1X ERGO objective allows
you to continuously change the eye level
through a 40 mm range while maintaining a
constant working distance (50 mm).
1 Rotate the upper knurled ring on the
Plan 1X ERGO objective to attach it to
the zooming body.
2 Rotate the focus knobs on the focusing
stand to raise or lower the microscope to
set to the height best suited for viewing.
3 Rotate the lower knurled ring to focus
on the specimen. Use the microscope as
usual after setting the eye level.
Note) When attaching to the SMZ1000, vignetting will occur in the lateral
direction of the viewfield at the high eye level with 1X or lower
zoom magnification.