Using Angle Beam Probes
3.2
When connecting an angle beam probe to the instrument, adjustments must be made for probe characteristics as well as test piece geometry.
These adjustments include:
•
Probe angle
•
Probe X value = the distance from the probe Beam Index Point (BIP) to the front edge of its wedge
•
Test piece thickness
•
O-Diameter
3.2.1
Setting the Angle Beam Probe Parameters
To configure the instrument for an angle-beam probe, follow this procedure:
In the SETUP menu, activate the EVAL submenu using the joystick (
1.
2.
Use the joystick (
performed by installing the correct probe/wedge and setting this value to > 89°.
Note: This function has both coarse and fine adjustment modes. Coarse adjustments are made with the function toggle, while fine
adjustments are made with the joystick.
USM Go Operator's Manual
) to select the function titled PROBE ANGLE. Input the angle for the installed probe. Surface evaluation is
).
Chapter 3. Making Measurements
73