Chapter 2. Instrument Setup

2.6.3 Checking the Calibration Results

Following the calibration procedure, the calculated acoustical velocity and probe delay are displayed. To view these calculated values:
1.
In the ACQUIRE menu, activate the RANGE submenu using the joystick (
2.
Press the function toggle or move the joystick (
PROBE DELAY - The adjustment made as a result of the AUTOCAL (zeroing) procedure is shown. This represents the time delay
caused by sound-wave traveling through the probe membrane, wear plate, or delay line.
VELOCITY - The calculated velocity after the calibration is shown. The material type is listed as "custom".