Chapter 2. Instrument Setup

2.4.2 Configuring the Instrument

Three instrument settings are directly dependent on the type of probe installed. These settings must be adjusted any time a probe of a
different type is installed, by following the instructions in the following sections.
2.4.2a
Selecting the Probe Type
In the ACQUIRE menu, activate the RECEIVER submenu using the joystick (
1.
2.
Use the joystick (
3.
To change the probe type, move the joystick (
the Icon Bar, near the upper left corner of the display, whenever that probe type is highlighted. The following options are available:
ON - Use for dual-element probes. If this option is selected, the Dual icon (see Figure 6 on page 17) is displayed.
OFF - Use for single-element probes. If this option is selected, the Single icon (see Figure 6 on page 17) is displayed.
4.
After completing your selection, move the joystick up or down (
38
(RECEIVER-DUAL)
) to select the function titled DUAL.
) or press the function toggle. Each available probe type is represented by an icon in
).
) to navigate away from this function.
USM Go Operator's Manual