Typical Calibration Setup (Pressure To Voltage); Zero Display Reading - GE Druck DPI 610 IS User Manual

Portable pressure calibrator/indicator
Hide thumbs Also See for Druck DPI 610 IS:
Table of Contents

Advertisement

Getting Started

Typical Calibration Setup (Pressure to Voltage)

Connect a device under test to the instrument as shown below.
Ext Press. Source
(Indicator Only)
Pressure
Regulator
General Procedure
Use the hand-pump to pressurise the system to the required level as
indicated on the display. Allow the display to settle and screw the volume
adjuster in or out as a fine adjustment to the required pressure. Record the
measurand (e.g.) Voltage, reading at each applied pressure.

Zero Display Reading

Both the input and output windows can be set to zero by operation of the ZERO
key, providing that the displayed reading is already within 5% of zero. To zero
either the INPUT (Measurand) or OUTPUT (Source) windows, proceed as follows
(e.g.),