ProCare 6000 Operation Manual
Canceling a Patient Call
All patient calls can be canceled by pressing the CANCEL button on the patient station.
Normal patient calls can also be canceled by registering staff presence within the pa-
tient's room. Normal patient calls can also be canceled by answering them at the base
master station. When a patient call is canceled, the following events occur:
The CALL light extinguishes.
•
The associated corridor light extinguishes.
•
The call disappears from the base master display.
•
The tones cease at the base master and at stations where the system monitor function is
•
active.
Canceling a Reminder
When a patient call is placed on reminder, the reminder can be canceled by pressing the
station's CANCEL button. It can also be canceled by registering staff presence in the
room (either manually by pressing the button on a presence station, or automatically by
wearing a staff IR locator badge). The registered presence must be the same level as the
set reminder.
When CANCEL is pressed or presence is registered, the following things happen:
The station's CALL LED extinguishes.
•
The associated corridor light extinguishes.
•
Note:
If presence is registered, the section of the corridor light that corresponds to the regis-
tered presence level stops flashing and steadily illuminates.
The reminder recall disappears from the base master display and from stations where
•
the system monitor function is active.
The CALL Light
4-7