Removing The Covers - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing the Covers

Removing the Covers
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
Removing the Outer Cover
This procedure is best performed with the analyzer resting on its front
CAUTION
handles in the vertical position. Do not place the analyzer on its front panel
without the handles. This will damage the front panel assemblies.
Refer to
Figure 7-1
for this procedure.
1. Disconnect the power cord.
2. With a T-20 TORX driver, remove the strap handles (item
(item
) on both ends until the handle is free of the analyzer.
3. With a T-20 TORX driver, remove the four rear panel feet (item
center screws (item
4. Slide the four bottom feet (item
5. Slide the cover off of the frame.
Removing the Inner Cover
Refer to
Figure 7-1
for this procedure.
1. With a T-10 TORX driver, remove the 11 screws (item
2. Lift off the cover.
Replacement Procedure
Reverse the order of the removal procedures.
7-6
).
) off the cover.
PNA Series Microwave Network Analyzers
) by loosening the screws
) by removing the
).
Service Guide E8361-90001
E8361A

Advertisement

Table of Contents
loading

Table of Contents