Zero Operation - AND Check Weighing Indicator AD-4404 Instruction Manual

Check weighing indicator
Table of Contents

Advertisement

8.16. Zero Operation

8.16.
8.16.
8.16.
The zero operation zeroes the display and changes to the gross display.
The operation can be performed from the front panel key or the input terminal of the
control I/O.
The adjustable range is based on the zero calibration and
The range is displayed in the unit of percentage of the weighing capacity.
If the operation can not be achieved, an error message and
The re-zero data is maintained in memory even without power.
Key Operation
Pressing and holding the SHIFT key, press the ZERO key.
Operation from Control I/O
1n f-xx
In
[
]
xx
01
Zero
[
]
xx is terminal number (01 to 11).
[Function] - [Set function] - [control I/O] - [Input]
Concerning Parameters
genf-1
Zero range
[
]
Range is within 30% of the weighing capacity.
[Function] - [Set function] - [General] - [Weight]
genf-9
Tare & zero unstable WGT
[
]
The prohibition of the re-zero operation in the unstable condition.
[Function] - [Set function] - [General] - [Weight]
genf-13
Clear mode at power ON
[
]
Specify power-on status.
[Function] - [Set function] - [General] - [Weight]
Zero Operation
Zero Operation
Zero Operation
(A
)
.................The input terminal of the control I/O.
xx
................................When it turns on, zero of gross is displayed.
Page 62
[genf- 6].
Zero range
ZR.ERR
is displayed.

Advertisement