3.3. Connecting the Loadcell Cable
3.3.
Connecting the Loadcell Cable
3.3.
3.3.
Connecting the Loadcell Cable
Connecting the Loadcell Cable
Caution
Do not share the loadcell cable with noise-generating devices or power lines,
because the loadcell signal is very sensitive.
We recommend that you use a 6 wire shielded cable to prevent loss of weighing
precision.
If the length of loadcell cable is shorter than 5 m, you may use a 4 wire shielded cable
with terminals 1 & 2 connected together (connected EXC+ and SEN+) and terminals
3 & 4 connected together (connected EXC- and SEN-).
Adaptable Compression Terminal Parts
Use the appropriate compression terminal parts to attach the cables.
Loadcell Output Adjustment for Zero Calibration (Zero Point Adjustment)
If the message "
If the message "
Use a resistor of more than 50 k with low (good) temperature coefficient, when
adding a resistor, to adjust the loadcell output, to the indicator terminals.
In Case of Positive Offset
CERR2
" is displayed, the zero point of zero calibration is too large.
CERR3
" is displayed, the zero point of zero calibration is too small.
In Case of Negative Offset
Page 14