Parameter List - AND Check Weighing Indicator AD-4404 Instruction Manual

Check weighing indicator
Table of Contents

Advertisement

11.3. Parameter List

11.3.
11.3.
11.3.
Category address: [Function] - [Set Function] - [General] - [Weighing]
Category
address
symbol
Display
genf- 1
refresh rate
Digital
genf- 2
filter
Sampling
genf- 3
freqeuncy
divider
Stability
genf- 4
detection
time
Stability
genf- 5
detection
motion
genf- 6
Zero range
Zero tracking
genf- 7
time
Zero tracking
genf- 8
motion
Tare & zero
genf- 9
unstable WGT
Tare at (-)
genf-10
gross weight
genf-11
Preset tare
AD-4404
Parameter List
Parameter List
Parameter List
Name
1: 5 times / second
2: 10 times /second
3: 20 times / second
0: Not used
1: 11 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
Use to decrease the cut-off frequency of
the digital filter.
The detection condition concerning
stablility.
0.0 : Stable at anytime.
The range to zero the gross display.
Center of range is zero calibration.
Unit: percentage of weighing capacity.
The function automatically traces the
weighing deviation at nearly zero point
and keeps zero display of gross display.
If parameter is 0.0, it does not work.
When unstable weighing, whether zero
or tare command is used.
0 :Disabled
1 :Enabled
When negative weghing, whether tare
command is used.
0 :Prohibit tare.
1 : Permission to tare.
Preset tare
0 :Not used
1 :Use
Descriptions
Select 3dB band for two
low pass filters.
A/D converter
Second filter
Relations:
genf-4, 5
Page 127
Range
and
choices
1 to 3
0 to 99
First filter
0 to 10
0.0 to
9.9 s
0.0 to
9.9 d
0 to 30 % 5 %
0.0 to
9.9 s
0.0 to
9.9 d
0 to 1
0 to 1
0 to 1
Default
2
4 8
1
1.0 s
2.0 d
0.0 s
0.0 d
1
1
1

Advertisement