A&D Data Format
A&D Data Format
9.2.6.
9.2.6.
A&D Data Format
A&D Data Format
This format is used for stream mode, auto print mode and total print.
This format is compatible with the AD-4325 indicator.
1
2
3
4
S
T
,
N
Header 1
Header 1
ST
Stable
US
Unstable
LO
Out of range
Header 2
GS
Gross value
NT
Net value
TR
Tare value
Weighing data
The data uses BCD code, is 7 figures and includes a decimal point.
When the data is negative, a minus sign is appended to the head.
When the data is out of range, all numerical characters are space (20h).
Unit
kg, g or t
Terminator
CR or CR + LF
5
6
7
8
T
,
+
0
Header 2
Weighing data
9
10
11
12
0
1
2
.
CR: 0Dh, LF: 0Ah