10.4. Initializing Parameters

10.4.
10.4.
10.4.
This function initializes the parameters stored in the indicator.
The parameters are stored in the flash memory and backup RAM.
Caution
There are reset functions that require re-calibration of the indicator
Note where the parameters are stored.
Kinds of intialization mode
The menu to initialize parameters is as follows:
[Function] - [Set function] - [SystemMgmt] - [init] - [RAM]
Display
RAM
Initializing RAM
Code
Initializing all codes
Initializing the
Function
function parameters
Initializing caribration
CAL
data
Initializing all
All
parameters
#1 If the switch on the A/D board set to "Disable", it can not be performed.
The Location of the Parameters and Objects of Initialization Mode
Backup RAM
Location
Flash memory
RAM
Code
Kinds
Function
CAL
All
#1 Code data can be stored into flash memory. Refer to " 8.26. Memory Backup"
#2 Total data is always stored in backup RAM.
AD-4404
Initializing Parameters
Initializing Parameters
Initializing Parameters
Kinds
The backup RAM is reset.
Zero point of the gross display, tare value zeroes.
All parameters of all codes are reset.
Resets parameters of the function list in flash memory.
Calibration data is reset in flash memory. If this
function is used, calibrate the indicator. #1
All parameters and calibration data are reset.
Recalibrate and input all parameters. #1
o
o
o
[Code]
[Function]
[CAL]
[All]
Description
o
o
o #1
o
o
o
o