Zero Calibration; Span Calibration - AND Weighing Indicator AD-4406 OP03 Instruction Manual

Table of Contents

Advertisement

5.2.8.

Zero Calibration

Step 8 Check the CAL 0 display.
Select a zero calibration method to adjust the zero
point.
Weighing input
(Normal way)
Digital input
Weighing Input
Step 9 Place nothing on the weighing unit. Press the ENTER
key after the STABLE mark has turned on. The new
zero point parameter will be stored. Proceed to step11.
ESC key ......... To proceed to step 11 without changing
Caution Do not press the
mark is off (detecting motion). Arrange the
condition using the F00 filter function.
Digital Input
Step10 Pressing the
of the zero point is displayed in the unit of mV/V. Adjust
the input voltage using the following keys.
key .......... To select a digit.
key .......... To select a value.
ENTER key .... To store the zero point parameter and
ESC key ......... To proceed to the next step without
5.2.9.

Span Calibration

Step11 Check the capacity display after CAL F is displayed for 2 seconds.
Select a span calibration method to adjust the capacity.
Weighing a mass less than
the maximum capacity
Weighing a maximum
capacity mass
Digital input
5.2. Calibration Procedure
The adjustment method
with nothing on the
weighing unit.
The numerical way to enter
a load cell output voltage.
the zero point parameter.
key while the STABLE
ENTER
key, a stored input voltage parameter
+/-
proceed to the next step.
changing the parameter.
The method to weigh a mass less
than the maximum capacity.
The method to weigh a mass
equivalent to the maximum capacity.
The numerical way to enter a load cell
output voltage.
To step 9
To step 10
To Span Calibration
Page 14
Zero calibration
Place nothing on
the weighing unit
(No load)
Turn on STABLE mark
F 2
ENTER
Stored input voltage
+/-
mV/V
ZERO
TARE
mV/V
New input voltage
F 2
ENTER
To step 12
To step 14
To step 16
AD-4406 Weighing Indicator