The Parameter List For The Function List; Basic Functions - AND AD-4430B Instruction Manual

Din rail weighing module
Hide thumbs Also See for AD-4430B:
Table of Contents

Advertisement

7.6.

The Parameter List For The Function List

When performing maintenance, use the following list as a memorandum.
When making inquiries about the product, inform your local A&D dealer of the user settings.
7.6.1.

Basic Functions

Function No.
Function
Parameter
name
and range
Fnc01
Disable key
0000 to 1111
switch
Fnc02
key
0 to 7
Fnc03
Display
1 to 3
update rate
Fnc04
x display
0 to 9
Fnc05
Digital filter
0 to 16
Fnc07
Hold
1 to 3
Fnc08
Set value of
-99999 to
near-zero
99999
AD-4430B
Description
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
0 : Enables
Key assignment to each binary digit.
1 : Disables
4th
0 : None
1 : Manual print command
2 : Hold
3 : Alternate switch
4 : Momentary switch
5 : Changing between gross weight and net weight
6 : Clear the tare weight
7 : Clear the zero value
NOTE: Clearing the zero value can be enabled
or disabled by CAL15.
1 : 20 times/second
2 : 10 times/second
3 : 5 times/second
0 : None
1 : Zero tracking in progress
2 : Alarm (Zero range setting error, over)
3 :
key status
4: : Zero band
5 : HI output (Over the upper limit value)
6 : OK output (Between upper and lower limit values)
7 : LO output (Below the lower limit value)
8 : User input 1
9 : User output 1
Selects a cutoff frequency.
0: None
1: 100.0 Hz
2: 70.0 Hz
3: 56.0 Hz
4: 40.0 Hz
5: 28.0 Hz
6: 20.0 Hz
7: 14.0 Hz
8: 10.0 Hz
1 : Normal hold
2 : Peak hold
3 : Averaging hold
The reference value for near-zero.
Decimal point position is linked to CAL02.
3rd
2nd
1st
9: 7.0 Hz
10: 5.6 Hz
11: 4.0 Hz
12: 2.8 Hz
13: 2.0 Hz
14: 1.4 Hz
15: 1.0 Hz
16: 0.7 Hz
- 43 -
Default
User
MEMO
value
setting
0000
(binary)
0
1
0
15
1
10

Advertisement

Table of Contents
loading

Table of Contents