11.2.9
Operating and Auxiliary Materials
11.3
Specifications of Parameters
11.3.1
Shaker Drive
29.11.2024
NOTICE
Using the wrong auxiliary materials can result in significant damage
to property.
Only use the auxiliary materials prescribed by the manufacturer in
accordance with the table below.
Description
Refrigerant (primary cooling
cycle, cooling compressor)
Cooling liquid (secondary
cooling cycle)
Detergent
Disinfectants
Data
Drive
Direction of rotation
Throw
Increment
Accuracy control
(at maximum rotation
speed, full scale)
Multitron Standard - Operating manual
Technical Data
Approved/applied products
R134a
Based on 1,2-propandiol with
inhibitor (must be suitable
for copper)
Approved for the food and
pharmaceutical sectors
Ex factory: 30 % Antifrogen L,
70 % water
Mild neutral cleaning agent
Dishwashing detergent
Quaternary ammonium com-
pounds
Value Unit
External rotor
 
motor
Clockwise  
25 or 50 mm
1 min
± 1
%
-1
Page 93 of 104