Multitron Standard - Operating manual
Setup and Function
3.3.2
Ethernet Interface
3.3.3
USB Connection
Page 28 of 104
The device has an Ethernet interface (RJ45 socket). This is located on
the right side of the device. The Ethernet interface can be used for the
following purposes:
Integration of the device into a network to control the device via
■
eve
the bioprocess platform software.
®
Control of the device via a laboratory management or monitoring
■
system from a third-party manufacturer. For this purpose, a driver
must be developed which translates the protocol used for the
external software. The necessary documentation for the communi-
cation protocol can be ordered from INFORS HT.
The following data and states are transmitted via the Ethernet inter-
face:
From the device to the SCADA/bioprocess software: setpoints and
■
actual values of the parameters as well as the status of the door
and PIN function
From the SCADA/bioprocess software to the device: sending new
■
setpoints
Alarms and error messages are not transmitted via the
Ethernet interface. Alarm limits can be defined either
in eve
or in the SCADA system, depending on the
®
application.
For more information on using the Ethernet interface, see ⮫ Chapter
6.5 'Using eve
to Operate the Device' on page 68.
®
The USB port is reserved for service purposes and has no function in
normal operation.