Replacing the Adhesive Matting
29.11.2024
Prior to the shaking process, gently pull on every vessel to ensure
■
they are all stuck on firmly.
If the humidity is too high or the temperatures are too low or if
■
there are extreme jumps in temperature (e.g. when using the timer
function), keep an eye on condensate. Condensate can result in
vessels detaching from the adhesive matting.
To remove vessels, gently and evenly pull or push on the neck and
■
wait a few seconds.
Never apply force!
In case of large vessels, it can take 20 to 30 seconds until they
■
detach from the adhesive matting.
Vessels that are stuck can be removed from the adhesive matting
■
by using a syringe to inject water under the flask.
Due to their shape (wide base, short neck), Fernbach flasks in
■
particular can be difficult to remove. If necessary, cover part of the
adhesive matting with the protective foil provided.
The adhesive power can deteriorate over time due to dust and dirt.
■
To clean and restore full adhesive power, thoroughly wipe down
the surfaces with a brush or a dish sponge and clear water with
mild detergent (dishwashing liquid). Then allow to dry overnight.
Quaternary ammonium compounds are suitable for disinfection.
■
Note the application time and rinse thoroughly with water after-
■
wards. If the adhesive matting is disinfected at regular intervals, it
might need to be replaced sooner.
To replace the adhesive matting proceed as follows:
1.
Thoroughly moisten the tray with water.
2.
Release the adhesive matting on one side of the tray and then
pull upwards at an angle.
3.
Degrease the tray with acetone and apply new wet adhesive
matting (as per separate installation instructions). Only remove
protective foil prior to use.
The removed adhesive matting can be reused and can be
reapplied after regeneration in water.
Multitron Standard - Operating manual
Accessories
Page 41 of 104