Multitron Standard - Operating manual
Setup and Function
Operation
RPM
Page 24 of 104
The load weights must be within the permissible
range. If the load is too light or too heavy, this will
prevent the table from running smoothly, leading to
increased wear on the bearings and joints.
The permissible load weights depend on the posi-
tion of the device in the stack, the throw and the
shaking speed (⮫ Chapter 11.3.3 'Ideal Loading Weights'
on page 96).
The shaker drive is operated using the "Rotation speed" parameter
(RPM, Rotations Per Minute). For an exact description of how to set
the parameters, see ⮫ Chapter 6.3 'Setting, Activating and Deacti-
vating Parameters' on page 60.