3.2
Working Light
3.3
Connections and Interfaces
3.3.1
Mains Connection
29.11.2024
The device features two LED spots to illuminate the incubation
chamber. These are located in the door area on the inside of the
casing.
The working light is switched on automatically in the following situa-
tions:
When the device is switched on using the power switch
■
When the door is opened or closed
■
When any of the buttons on the operating panel are pushed
■
The working light remains switched on for two minutes, then is
switched off automatically.
The behaviour of the working light described above
represents the standard configuration. If required, the
working light can be configured to be constantly
on or off. The working light setting is made in the
service menu and can therefore only be changed by
an INFORS HT service technician or licensed dealer.
The mains connection is located on the right-hand side of the device.
Three different versions of the device are available for different mains
voltages:
230 V 50 Hz
■
230 V 60 Hz
■
115 V 60 Hz
■
Two device fuses immediately adjacent to the mains connection pro-
tect the device from impermissibly high power input.
The country-specific power cable required for connecting the power
supply is included in the device's scope of delivery. If the power cable
is defective, replace it with a power cable of the same type.
Prior to connecting the device, make sure that the voltage values of
the device match those of the local mains voltage. The mains connec-
tion must be easily accessible at all times so that the device can be
disconnected from the power supply quickly in case of an emergency.
For information on the electrical connection values, see ⮫ Chapter
11.2.3 ' Electrical Connection and Power Values' on page 90.
Multitron Standard - Operating manual
Setup and Function
Page 27 of 104