Infors HT MULTITRON STANDARD Operating Manual page 54

Incubator shaker
Hide thumbs Also See for MULTITRON STANDARD:

Advertisement

Multitron Standard - Operating manual
Operation
6.2.1
Opening the Door
Page 54 of 104
The door can be opened while the device is running. The table and
temperature control are stopped as soon as the door is opened more
than 30°. It is not possible to open the door fully until the table has
come to a full stop. The table will start up again automatically as soon
as the door is closed more than 45°.
CAUTION
If the door is opened while the shaker drive is running, the table
can cause injury due to its significant momentum. As such, take care
when opening the door while the shaker drive is running, and wait
until the table has come to a complete standstill.
NOTICE
The door can only be opened fully when the device is switched on
at the power switch and there is no load on the half-open door.
If the door is pushed down violently, this can damage the door
mechanism.
Under no circumstances leave the door hanging halfway in the lock.
Support the door by hand until the table has come to a complete
stop and the door lock is heard to unlock.
In order to open the door while the shaker drive is running, proceed as
follows:
1.
Temporarily stop the device by slightly opening the door.
2.
Wait until the table has come to a standstill.
�� The door is unlocked and remains unlocked for five seconds.
3.
Open the device door all the way. If the door does not unlock
straight away, do not lean on it; instead, lift it up slightly to take
the load off the locking mechanism. If the door is not opened
within five seconds, it is locked again and must be fully closed
and opened again.