6.8
Behavior in Case of a Power Interruption
29.11.2024
To switch off the device, proceed as follows:
1.
Deactivate all active parameters.
2.
Press the power switch on the top left of the casing.
3.
If the device is not going to be used for an extended period of
time, unplug the power plug.
If the power supply to the device is interrupted during a running culti-
vation process (e.g. by flicking the power switch or in case of a power
failure), all parameters and timer setpoints as well as the residual time
of the last active timer phase are stored.
If the power supply is restored, the device automatically restarts with
the last stored setpoints. If a timer was active prior to the interruption
to power supply, the device restarts with the residual time of the last
active phase and the setpoints stored for this phase.
P. out appears in the view boxes as a warning. The
The alarm message
alarm message can be confirmed by pressing any button; and then
disappears.