7.3
Fault Tables
General Faults
Fault description
After activating the
power switch, the dis-
play does not light up.
Door does not open
fully.
The tray is not released
when the door is
opened.
The tray cannot be
locked.
The working light does
not work.
29.11.2024
The tables below describe potential faults that do not usually trigger
an error message on the display or – with a few exceptions – an alarm
signal.
Cause
Power supply of the device
interrupted.
The device fuse has triggered.
The device is switched off.
The door is under pressure.
Door mechanism defective.
The door is not fully open.
There is a foreign object jam-
ming the locking mechanism.
The tray is jammed due to for-
eign objects under the table.
The tray is bent.
The working light is deacti-
vated in the service mode.
Multitron Standard - Operating manual
Remedy
Check if the plugs are
■
plugged in correctly.
Check the mains con-
■
nection.
Change the fuse (⮫ chapter
7.4, page 78). If the fuses
trigger several times, con-
tact your local INFORS HT
representative.
Switch on the device.
Close the door completely
and then open it again.
Contact your local INFORS
HT representative.
Push the door down to its
bottom position.
Remove the foreign object.
Detach the table from
its axis, tip it up and
remove the foreign objects
(⮫ chapter 8.2.3, page 82).
Replace the tray.
Have the working light acti-
vated. Therefore contact
your local INFORS HT repre-
sentative.
Rectifying Faults
Personnel
Operator
Operator
Operator
Operator
INFORS HT service
technician or
licensed dealer
Operator
Operator
Operator
Operator
INFORS HT service
technician or
licensed dealer
Page 75 of 104