6
Operation
6.1
Switching the Device on
6.2
Loading the Device
29.11.2024
NOTICE
The shaker drive can start up automatically, if the device was not
switched off correctly beforehand. Lose items in the incubation
chamber can damage the device and cultivation vessels if this hap-
pens.
Activate the power switch on the right side of the housing.
�� Once it has been switched on, the device will automatically
run a self-test. The view boxes display the message
If the device was not switched off correctly (parameters still
activated), the message
can be cleared by clicking any button.
�� Once the initialisation process is complete, the
on/off function symbol will light up briefly. The device is then
ready for operation.
CAUTION
If a vessel breaks at high rotation speed, the glass splinters are slung
around the interior. When the door is open, the device does not halt
immediately so pieces of glass can be slung out.
Do not open the door in a panic if you notice that a vessel has
broken.
Switch off the device at the power switch and only open the
door when the table stands still.
CAUTION
If you put a load of more than 20 kg on the door of the device, there
is a risk that the door might break. This can lead to property damage
and injuries.
Do not lean on the door.
Do not put a load of more than 20 kg on the door.
Multitron Standard - Operating manual
Operation
P. out then appears. This message
Power supply
HELLO .
Page 53 of 104