Multitron Standard - Operating manual
Installation and Commissioning
Device under a Bench
100
100
Page 52 of 104
When placing the device under a bench, observe the following:
You must leave a ventilation gap of at least 100 mm between the
device and rear wall of the bench to ensure that heat produced by
the device can escape.
The minimum distance between the device and the underside of
the bench top is 100 mm.
NOTICE
Failure to comply with the required minimum distances can result
in components of the device overheating and the device being
damaged. As such, please observe the following points:
Never cover the air vents on the right-hand side or those on the
rear of the device.
Never position the device right up against a wall.
29.11.2024