Fault description
The temperature fails to
reach the desired set-
point.
Poor cooling perform-
ance. The desired set-
point cannot be reached
even though the cooling
unit is in operation.
29.11.2024
Cause
Error during temperature measure-
ment.
Fans are defective, air circulation in
the incubation chamber is therefore
insufficient.
The door is not closed completely.
The room temperature is too high.
IMPORTANT: The room tempera-
ture refers to the temperature
directly on the device. This can be
significantly higher than the tem-
perature in other parts of the room.
Other devices with strong heat radi-
ation are located in the immediate
vicinity of the device (e.g. ultra-low
freezers or refrigerated centrifuges).
A barrier in the room prevents cir-
culation of cold air.
The room temperature is not con-
stant (e.g. because the air condi-
tioning is turned down on the
weekend).
Required minimum distances for
air circulation are not observed.
The device has been pushed back
against the wall.
Multitron Standard - Operating manual
Rectifying Faults
Remedy
case of a defect, con-
tact your local INFORS
HT representative.
Contact your local
INFORS HT represen-
tative.
Close door com-
pletely.
1.
Reduce the
room tempera-
ture.
2.
Use a fan to
improve air cir-
culation.
3.
Move the
device.
1.
Use a fan to
improve air cir-
culation.
2.
Place a barrier
between the
devices.
3.
Move the
device.
1.
Remove the
barrier.
2.
Move the
device.
Ensure that the room
temperature is con-
stant.
Move the device to
ensure that heat can
escape and heat does
not accumulate.
Personnel
INFORS HT service
technician or
licensed dealer
Operator
Operator
INFORS HT service
technician or
licensed dealer
Operator
INFORS HT service
technician or
licensed dealer
Operator
INFORS HT service
technician or
licensed dealer
Operator
INFORS HT service
technician or
licensed dealer
Page 77 of 104