7    Troubleshooting; Performance Troubleshooting - Invacare Modulite User Manual

Hide thumbs Also See for Modulite:
Table of Contents

Advertisement

7    Troubleshooting
7.1    Performance Troubleshooting
For additional troubleshooting information regarding the power wheelchair & electronics, refer to the Troubleshooting section of the 
power wheelchair and remote user manuals (provided separately).
Symptom
Wheelchair power is ON, but system does not 
drive
Seating system not functioning
1656515-G
Probable Cause
System tilted and / or elevated beyond 
the drive lockout (DLO) angle
Drive motors not engaged
LNX powered central legrest with 
telescoping footboard lowered
Low batteries
Loose/faulty electrical connection
Blown fuse
Interference / obstructions, pinched 
wires
Solutions
Return seating system to neutral (home) 
position.
Engage drive motors.
Retract footboard to top position.
Check / charge / replace batteries.
Contact your provider.
Check cable connections / check cable ties (too 
tight / too loose).
Contact your provider.
Inspect / replace fuse.
Contact your provider.
Check for sources of interference or 
obstructions / inspect cables for pinch points.
Contact your provider.
7    Troubleshooting
56

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Modulite and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents