Table 6-1. Aim/Aim Plus Alert Messages - Abbott aim Technical & Service Manual

Table of Contents

Advertisement

Section 6, Troubleshooting
Table 6-1. AIM/AIM Plus Alert Messages
Message
Condition
Remedy
PROGRAMMING
Programming is not complete and no | Press any key to clear.
INCOMPLETE
keys have been pressed for more
than three minutes.
START
Programmed pump has remained in | Press [SILENCE] to mute for three
the stop mode for more than three
minutes.
minutes.
Press [START] to begin protocol or
move ON/OFF switch to OFF position
fo turn pump off.
CHECK PRINTER
Pump to printer connection has been | Press [SILENCE] key to mute for two
disrupted or print function has been
| minutes.
selected without the printer
connected.
Check printer cable connections to
pump and printer.
Check printer for malfunctions
according to manufacturer's
recommendations.
ALMOST EMPTY
Delivery will complete in 30 minutes
| Press [SILENCE] to mute for ten
or less.
minutes.
For bolus only program, delivery will
| Note: This alert does not occur with
complete during next bolus delivery. | TPN, Intermittent, or Variable Time
protocols,
EMPTY CONTAINER
Pump has delivered container size
Press [SILENCE] to mute for two
programmed.
If bolus delivery is in progress when
alarm occurs, bolus delivery can be
completed or deleted as desired after
New Container is chosen.
minutes.
Press [STOP].
Use New Container function to repeat
current program; refer to the System
Operating Manual for
instructions.
END OF INFUSION
For TPN protocols: Pump has
delivered TPN volume or completed
auto taper down.
For Intermittent protocols: Pump has
reached end of off cycle of last dose
delivered.
Press [SILENCE] to mute for two
minutes (periodic beeping during
KVO delivery cannot be muted).
Press [STOP] to place pump in stop
mode.
AIM/AIM Plus Technical Service Manual
6-3
430-600035-002 (Rev. 11/99)