Section 5, Maintenance and Service Tests
3.
Connect the 12V power supply to the turns counter. Press RST on the turns counter to reset the
display to 0.0.
Wet Test Setup
1.
Install a primed cartridge set into the pump. If necessary, refer to the pump's System Operating
Manual for detailed instructions regarding cartridge installation and priming.
2..
Position the proximal end of the tubing in the beaker of water and the distal end in a 20 mL
graduated cylinder to measure the liquid delivered.
5.3.3.3
Test Procedure
0.01,
INFUSED
PRESS
START
TO
INFUSE
DELIVER
THE
LOADING
DOSE
NOW?
YES
OR
NO
#.#mL
INFUSED
DELIVERING
LOADING
DOSE
1]
1 |]
AIM/AIM Plus Technical Service Manual
In the stop mode, press [START].
Press [YES] to begin delivery of the Loading Dose.
Verify that the motor begins turning clockwise and that the volume
infused accrues on the screen (delivery rate 125 mL/hr).
The pump requires approximately five minutes to deliver the 10
mL dose or ten minutes to deliver the 20 mL dose. When the
loading dose delivery is complete, the pump returns to the stop
mode.
For Dry Test:
Verify that the display on the turns counter is
between 9.95 and 10.05 mL. If using the PVT Form, record the
results. Disconnect the turns counter from the pump.
For Wet Test: Verify that the total amount of fluid in the container
is between 19.0 and 21.0 mL. If using the PVT Form, record the
results,
5-23
430-600035-002 (Rev. 11/99)