Abbott aim Technical & Service Manual page 65

Table of Contents

Advertisement

5.3.8
PVT Data Form
Section 5, Maintenance and Service Tests
D AIM List 13038-04
CO AIM Plus List 13967-04 | Serial
No:
Performance Verification Test
Value
Pass | Fail | N/A
Electrical Test
Set Clock
Software Version Verification
Main/Safety
Supply Check
EXT AC (7.60 - 8.40V)
REG SV (4.80 - 5.20V)
Display Test
Contrast Adjust
LED Test
Keypad Test
Battery Check
Display Test
ALK (4.60 - 5.16V)
LITH (2.79 - 3.50V)
Idle Current
DMM (30 mA or less)
Optics Check |
AIR (231 - 255)
OCCL (231 - 255)
Optics Check II
AIR (14 to 40)
OCCL (12 to 40)
Clearing Memory
Beeper Test
Duty Cycle (approx. 50%)
Frequency (AIM 4 kHz, AIM Plus 2.00 - 2.30 kHz)
Motor Test
Dry Test or Wet Test (circle one)
DMM (210 mA or less)
1115 or less)
RPM (76.0 - 97.0)
Servo Calibration
Dry Test or Wet Test (circle one) | :
MD (50 - 75)
LR (>25)
Dry Test: ILO (200 - 450)
Wet Test: ILO (100 - 450)
Dry Test: IHi (275
- 600)