Abbott aim Technical & Service Manual page 75

Table of Contents

Advertisement

Section 6, Troubleshooting
Table 6-3. AIM/AIM Plus Alarm Codes 008 - 143
Code
Cause
Troubleshooting (repair as required}
107-AC POWER
AC adapter
Power off the pump. Verify use of manufacturer's
supplying more than | recommended AC power supply. Use a multimeter to
9.0 volts.
verify power supply voltage. Replace power supply as
required. Power on again with correct power supply or
batteries only.
108-MAIN ROM CKS | Main
Power off, wait one minute, then power on again to clear
microprocessor
alarm. If alarm does not reoccur, it has been cleared.
6811 - ROM
]
checksum error.
If alarm reoccurs, power off the pump and access the test
-
mode as described in Section 5.3.2.3, Entering the
109-MAIN RAM TST | Main
Service Test Mode. Select 02 CLEAR EXT RAM
microprocessor
then 03 CLEAR ERR LOG. Power off, wait one
6811 - RAM error.
minute, then power on again. If alarm does not
110-MAIN STACK
Main
reoccur, it has been cleared.
6814 stack
| If alarm reoccurs, check CPU U19.
overfiow.
For code 108, check EPROM U15.
111-MAIN CPU TST | Main
Replace the CPU U19.
microprocessor
6811 - background
CPU test failed,
112-NO MOTOR CTL | Motor power up test | Check motor or motor drive circuit, encoder or encoder
failure.
interface circuit.
113-NO RAM TEST | RAM check did not
| Power off, wait one minute, then power on again to clear
complete.
alarm. If alarm does not reoccur, it has been cleared.
If alarm reoccurs, power off the pump and access the test
mode as described in Section 5.3.2.3, Entering the
Service Test Mode. Select 02 CLEAR EXT RAM
then 03 CLEAR ERR LOG. Power off, wait one
minute, then power on again. If alarm does not
reoccur, it has been cleared.
If alarm reoccurs, check CPU U19 or RAM U14.
AIM/AIM Plus Technical Service Manual
6-9
430-600035-002 (Rev. 11/99)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aim plus13038-0413967-04

Table of Contents