AIM/AIM Plus Technical Service Manual
Section 5, Maintenance and Service Tests
Verify the AIR reading is 14 - 40
Verify the OCCL reading is 12 - 40.
Press [NO] to exit.
Remove the test block.
If using the PVT Form, record the results.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP] to repeat the test.
The pump clears the external RAM chip. As the test proceeds, the
address changes.
No user response required; pump automatically advances.
The pump clears the error logs. As the test proceeds, the address
changes.
No user response required; pump automatically advances.
The pump clears the NVRAM
chip. As the test proceeds, the
address changes.
No user response required; pump automatically advances.
5-15