Section 6, Troubleshooting
6.3.2
Alarm Codes
Following are descriptions, causes, and remedies of alarm conditions. Press the [HELP] key for
information about a current alarm condition. As necessary, use the arrow keys to scroll through the
message. Press the [BACK-UP] key to exit the help screen.
Before servicing the pump, attempt noninvasive recovery steps. If the alarm is not cleared, repair the
pump as needed, and perform the PVT before returning the pump to use (refer to Section 5.3,
Performance Verification Test).
The alarm log and pump software versions can be reviewed and printed through the Alarm Log
feature of the Options Menu (refer to the System Operating Manual for instructions).
Note: Error code 999 is displayed if multiple errors occur in one minute.
ERROR
DETECTED
999-HIGH
ERR
CNT
CONTACT
CUSTOMER
SUPPORT
Use the alarm log to itemize error numbers and refer to the
following alarm code tables for troubleshooting information.
WARNING:
Always disconnect the pump. from the patient before servicing the pump.
430-600035-002 (Rev. 11/99)
6-4
AIM/AIM Plus Technical Service Manual