This page contains information about the software installed on the Console.
4) CHARACTER SET (FOR SERVICE PURPOSE ONLY)
This page displays all the possible characters that the Console display can produce.
To exit from the Console BIOS, turn the System OFF using the main power switch on the
rear panel of the Console.

7.9 Setting the Console Max Flow Alert

Main Menu
Minimum Flow Alert
Maximum Flow Alert
Pressure Menu
Stopwatch*
* Monitor only
Extended Menu
To change the MAXIMUM FLOW ALERT setting, depress the MENU keypad until the
MAX FLOW ALERT option is displayed. Use the DECREASE or INCREASE arrow keys
to decrease or increase the alert setting. The default MAX FLOW ALERT value is set to
8.0 LPM at startup. The MAX FLOW ALERT threshold setting is continuously displayed in
the upper part of the display.
If the System detects flow in excess of 10 LPM, the digital flow display will not display a
numerical flow, but will display upwards pointing arrows: "^^.^^ LPM"
TM
CentriMag
Circulatory Support System Operation Manual (US)
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Figure 36: Monitor MENU Structure
Setting Maximum Flow Alert
Set RPM
Set RPM
Exit
Set RPM
Decrease
Set RPM
Decrease
Set RPM
[Start|Stop|Reset]
Set RPM
Stop Watch 2
Set RPM
Increase
Increase
Enter
[Start|Stop|Reset]
Stop Watch 1
Enter
Page 66 of 126