Section 5, Maintenance and Service Tests
5.3.4.3
Test Procedure
0.0mL
INFUSED
RATE
25.0mL/hr
PRESS
START
TO
INFUSE
#.#mL
INFUSED
RATE
25.0mL/hr
OCCLUSTON
#.#mL
INFUSED
DELIVERING
LOADING
DOSE
AIR
IN
LINE
#.#mL
INFUSED
RATE
25.0mL/hr
PRESS
START
TO
INFUSE
1 REVIEW
PROGRAM
2
HISTORY
3 KEYPAD
LOCK
USE
FOR
MORE
AIR
SENSITIVITY
1
HIGH
2
LOW
3 AIR
ALARM
OFF
USING
AN
AIR
ELIMINATING
FILTER?
YES
OR
NO
430-600035-002 (Rev. 11/99)
Tn the stop mode, press [START].
Wait for the alarm to occur.
Verify that OCCLUSION flashes on the display, the alarm LED
flashes and the alarm sounds. If using the PVT Form, record the
results,
Press [SILENCE] to quiet the alarm. Press [STOP] to return to the
stop mode.
Clear the occlusion by removing the clamp from tubing. Remove
the cartridge set from the pump and clear fluid from the tubing.
Install the dry set in the pump.
Press [START].
Wait for:'the alarm:to occur.
Verify that AIR IN LINE flashes on the display, the alam LED
flashes and the alarm sounds. If using the PVT Form, record the
results.
Press [SILENCE] to quiet the alarm. Press [STOP] to return to the
stop mode.
Press [OPTIONS].
Press [4] to select ATR SENSOR (the line does not need to be on the
screen to be chosen).
Press [3] to turn the air alarm off.
Press [YES].
5-26
AIMIAIM Plus Technical Service Manual